3 General settings for exploratory and determination mode
757 VA Computrace – Software
30
RDE/SSE
RDE stands for Rotating Disk Electrode and is used for direct
and stripping determinations with Solid State Electrodes (SSE).
Stirrer/RDE (rpm)
[ 0...3000 rpm ; 2000 rpm ]
Revolutions per minute of the rotating disk electrode. The
stirring of the RDE remains active during all preparation pro-
cedure steps until the start of sweep.
t
Surface
Notes:
The RDE can be used for all measurement modes.
For the 757 VA Computrace Stand, a drive shaft with different
electrode tips is available as an option (see Hardware Man-
ual).
For installation and maintenance of the RDE, see Hardware
Manual.