4.3 System functions
IC Net 2.1
25
Ident
User defined identifier (title) for the chroma-
togram to be displayed in the title bar of the
chromatogram window and in the
Chroma-
togram
open
window.
Calibration level
Calibration level (0 = sample; 1}n = cali-
bration solutions).
Info 1 / Info 2
Sample description.
Volume
Injected volume in PL.
Dilution
Dilution of the sample.
Vial number
Autosampler vial position to take sample
from.
Amount
Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:
C
s
= C
c
Amount
s
/
Amount
c
Internal standard amount
Concentration of the internal standard
component for relative concentration calcu-
lations.
Date/time when...
Date and time of sample collection (the de-
fault values are equal to the date and time
when the chromatogram starts).
The
Verify sample
option is disabled if determinations are made
with an active sample queue.