6 Devices
IC Net 2.3
92
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
Hardware error
Hardware error detected at the 761 Com-
pact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Alarm leak detector
Leak detector has detected solvent in the
instruments interior.
Remote line
Set remote output lines 18.
Selection: 0 (line off, inactive, open)
1 (line on, active, 0 V)
p (pulse output)
* (leave line in current status)
PowerOn values Display of the power on startup values for
remote output lines set on the Hardware tab.
Pulse length Length of a pulse in ms.
Alarm stops
The Alarm stops tab of the hardware settings window defines the
events for which the 761 Compact IC is stopped immediately. At
an alarm stop, high-pressure pump and peristaltic pump are
stopped immediately, the running determination and the active
sample queue are also stopped.