6.12 Metrohm 709 Solvent Delivery Unit
IC Net 2.3
135
Manual
The Manual tab of the Pump Control Unit window is only available
for a connected system.
Pressure Live display of current pressure in MPa.
This field is also available for the WATCH
WINDOW
.
Flow Live display of current flow in mL/min. This
field is also available for the WATCH WIN-
DOW
.
New flow A new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range:
0.01 ... 20.00 mL/min (depending
on number of pumps)
Min/Max pressure The minimum and maximum pressure limit
for the individual pump units can be set in
steps of 0.1 MPa.
Entry range: 0.0 ... 50.0 MPa.
%B Individual flow rate for pump B in percent of
total flow.
Entry range:
0 ... 100 %
%C Individual flow rate for pump C in percent of
total flow.
Entry range: 0 ... 100 %
%D Individual flow rate for pump D in percent of
total flow.
Entry range: 0 ... 100 %