8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
237
Checking Use for predose for Dosino 1, VMS is added to the
measuring vessel before the sample transfer. Then Suppressor
standard solution is added via Dosino 3. That should be specified
on the
Substances tab of the EDIT WORKING METHOD PARAME-
TERS
window (here the window for CVS):
Determination of the sample:
Choose "DT Suppressors with calibration curve" for Calibration
technique.
Settings in the DOSINOS window:
Checking Use for predose for Dosino 1, VMS is added to the
measuring vessel before the sample transfer. Then sample is
added via Dosino 3. That should be specified on the Substances
tab of the EDIT WORKING METHOD PARAMETERS window (here
the window for CVS):